The FCC-TEM-JM7D extends the operating frequency range beyond that of existing devices to now permit conducting emissions and immunity testing of Integrated Circuits to 3 GHz. The IC is mounted on a test fixture that bolts in place. Once installed, measurements as per IEC 61967-2 and -8, SAE 1752-3, and IEC 62132-8 can be performed to 3 GHz. It is configured to accommodate the standard 100mm x 100mm test plate specified in IEC 61967-2.

The FCC-TEM-JM7D can be used for immunity testing and requires less than 1 milli-watt input power to achieve a 10 volt per meter electric field and approximately 8 watts to achieve 1000 volt per meter electric field.

Product Specifications

Product
Frequency
Maximum EUT dimensions (mm)
Maximum Input Power (watts)
Maximum VSWR
Overall Dimensions (mm)
RF connector
FCC-TEM-JM7D DC - 3 GHz 50 x 50 x 8 250 1.2:1 274 x 127 x 61 SMA
Product specifications are subject to change without notice